Design of a Flat Panel Detector for the PET-Laparoscope System Using Simulations

Xiaodong Yue, Madhushanka Rukshani Liyanaarachchi, Kenji Shimazoe, Kenichiro Ogane, Hiroyuki Takahashi, Miwako Takahashi, Kazuaki Hara, Etsuko Kobayashi, Ichiro Sakuma

Journal Journal of Medical Robotics Research , vol. 7(01) , pp. 2241001 · 2022 · Cited by 0

Medical Instrumentation and DevicesBiomedical Imaging and Reconstruction

Abstract

To improve spatial resolution in the PET-laparoscope system for intraoperative metastatic lymph nodes detection, design of a flat panel detector was optimized using Monte Carlo simulation tools. The performance of the new fixed detector was evaluated in point radio-source and phantom experiments. The experiment result shows that the system with the new fixed detector design has a spatial resolution better than 10 mm in three dimensions, and the system can distinguish two radio-sources clearly in a 2-minute scan.